INSPECTION REPORT
Health Protection
PKUR-CM5PKU
PREMISES NAME
Fountain Wellness
Tel: (778) 590-1030
Fax:
PREMISES ADDRESS
102 - 7511 120th St
Delta, BC V4C 0C1
INSPECTION DATE
December 14, 2022
TIME SPENT
1.5 hours
OPERATOR (Person in Charge)
Noya Madadi
NEXT INSPECTION DATE
December 30, 2022
INSPECTION TYPE
Routine
ACTION TAKENPSE SERVICE TYPES
ADMINISTRATIVE
Information Provided
ENFORCEMENT
Require Corrections
Esthetics, Laser Services
HAZARD RATING FOR YOUR FACILITY: Low (≤15) Total Score = 15
Critical Hazards: Total Number: 1
313 - Adequate protective eyewear not available for operator and/or clients
Observation: Er: YAG function on IPL machine emits a wavelength of 2940 nm. Eyewear is labelled for 190-1200 nm only.
Corrective Action(s): Please obtain eyewear that covers the same wavlength that the machine emits Ensure it is OD 5+.
Violation Score: 15

Non-Critical Hazards: There are no non-critical hazards.
Follow Up to "Critical" Violations Noted on Previous Inspections (if applicable): No corrections entered
Comments

Routine inspection:

Personal services offered by aesthetician: chemical peels, microneedling, and laser hair removal (Er:YAG and Nd:YAG).
Services provided by a Registered Massage Therapist and services regulated by the Health Professions Act were not inspected.
Hand sink in treatment room provided with liquid soap and single use paper towels.

Chemical peels:
All lotions, chemicals, serums, etc dispensed from pump style bottles or into bowls.
Brushes are disinfected with 70% alcohol.
Microdermabrasion- operator states this service is not currently provided. Ensure a high-level disinfectant is obtained to disinfect microdermbrasion tips if this service resumes.

Microneedling:
All needles with cartridges received sterilized and packaged. Single use and discarded into sharps container.
Microneedling pen is covered with a plastic covering when in use.

Laser hair removal:
Laser warning sign posted on door.
No mirrors in treatment room.
Laser machines are key operated with emergency on/off buttons.
IPL laser machine hand piece that comes into contact with intact skin is disinfected with Caviwipes.
Nd:YAG function on machine emits a wavelength of 1064 nm. Protective eyewear is labelled for 190-1200 nm. Ensure it is OD7+ as per manufacturer's recommendations.
***Do not provide laser services until it is verified with the supplier that the protective eye wear is adequate.

Fat Cavitation Ultrasound:
Please provide the following as soon as possible:
1. Is the KIM 8 Slimming System Device considered a medical device and does it require a Health Canada Medical Device Active Licence?
2. Proof that the device is safe to use.
3. Information of any safety precautions for client and operator.
4. Information on how the device works.
5. Method of cleaning and disinfecting (as part of the infection control plan).
Note: As this machine as electrical supply, it must have either CSA or UL certification.

Please provide an infection control plan and training certificates by December 30, 2022.